Measuring the Zeta Potential of Highly Conductive Colloidal Samples

Diffusion barrier technique is a method of measuring zeta potential using Zetasizer Nano instruments, while preventing the denaturation of proteins.

This method also measures colloidal sample with high conductivity. However, an increase in conductivity affects the zeta potential measurements in such samples owing to a number of factors such as electrode polarization, sample degradation, electrode blackening and Joule heating.

Although the automatic settings of the instrument minimize these problems, the application of diffusion barrier technique will further reduce the effects of field on the zeta potential results.

This article describes the application of the diffusion barrier technique to samples with high conductivity by measuring various liposome samples.

Sample Preparation Method

Liposomes were prepared using sonication technique. Following this, dipalmitoylphosphatidylglycerol (DPPG) and dipalmitoylphosphatidylcholine (DPPC) were mixed in phosphate buffered saline (PBS) to prepare three different types of anionic liposomes. Table 1 shows the final concentration of the prepared liposomes.

Table 1. Summary of the liposome compositions prepared.

Liposome Composition DPPC (mg) DPPG (mg) PBS (mL)
1 29 1 3
2 27 3 3
3 25 5 3

Characterization of Liposomes

A Zetasizer Nano ZSP operating at 25°C was used to carry out both zeta potential and size measurements of the samples. For sizing measurements, the samples were diluted 1 in 100 with PBS using dynamic light scattering (DLS) technology. However, for zeta potential measurements, the measurement zone was inserted with a folded capillary cell filled with PBS and a 50µL aliquot of liposome sample via a gel loading pipette tip as shown in Figure 1.

Photograph showing a liposome aliquot injected into the measurement zone of a folded capillary cell containing PBS.

Figure 1. Photograph showing a liposome aliquot injected into the measurement zone of a folded capillary cell containing PBS.

Experimental Results

Measurement of zeta potential of samples prepared using PBS is quite challenging owing to a number of factors as discussed above. The z- average diameters and zeta potential means of three different liposome compositions were outlined in the Table 2. The results obtained are the mean values of five repeat measurements along with the standard deviations.

Table 2. The z-average diameters (in nm) and mean zeta potential values (in mV) of the various anionic liposomes prepared in PBS. The standard deviation values from five repeat measurements are shown in brackets.

Liposome Composition z-Average Diameter (nm) Zeta Potential Mean (mV)
1 133.5 (1.35) -5.8 (0.47)
2 77.4 (0.86) -14.4 (0.43)
3 67.1 (0.7) -19.2 (0.74)

The results showed excellent consistency during repeat measurements. In addition, the expected trend of the mean showed more negative zeta potential results with an increasing DPPG content.

Conclusion

The diffusion barrier method is a measurement method that minimizes the effects of field application over zeta potential results. This method was originally developed for measuring the protein electrophoretic mobility. However, it can also be used for the measurement of any sample prepared using medium with high conductivity, which is evident from the results of repeat zeta potential measurements in this experiment. Another major advantage of this technique is its low sample volume requirement i.e. below 20µL.

This information has been sourced, reviewed and adapted from materials provided by Malvern Panalytical.

For more information please visit Malvern Panalytical.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Malvern Panalytical. (2024, March 01). Measuring the Zeta Potential of Highly Conductive Colloidal Samples. AZoNano. Retrieved on November 21, 2024 from https://www.azonano.com/article.aspx?ArticleID=3921.

  • MLA

    Malvern Panalytical. "Measuring the Zeta Potential of Highly Conductive Colloidal Samples". AZoNano. 21 November 2024. <https://www.azonano.com/article.aspx?ArticleID=3921>.

  • Chicago

    Malvern Panalytical. "Measuring the Zeta Potential of Highly Conductive Colloidal Samples". AZoNano. https://www.azonano.com/article.aspx?ArticleID=3921. (accessed November 21, 2024).

  • Harvard

    Malvern Panalytical. 2024. Measuring the Zeta Potential of Highly Conductive Colloidal Samples. AZoNano, viewed 21 November 2024, https://www.azonano.com/article.aspx?ArticleID=3921.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.